Honeywell UDA2182 Bedienungsanleitung Seite 81

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 258
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 80
Configuration
January 2009 UDA2182 Universal Dual Analyzer Product Manual 69
Sub-menu
selection
Parameter Selection or
Range of
Setting
Parameter Definition
TDS Factor
(only PV Type
TDS)
0.010
1.000(default)
2.000
The TDS Factor is a conversion value applied to
conductivity to derive total dissolved solids, in units of
ppm per μS/cm.
Temp Type
8550Ω Therm
(default)
1000Ω RTD
Manual
8550Ω Thermistor
1000Ω Resistance Temperature Detector
Manual
Temp Deg C or F
-10.0 to
140.0ºC
14 to 284ºF
If “Manual” is selected at “Temp Type” -Temp Deg F
or C will appear depending on what Temperature Unit
was selected in “Maintenance” setup group,
parameter “Temp Units”.
Seitenansicht 80
1 2 ... 76 77 78 79 80 81 82 83 84 85 86 ... 257 258

Kommentare zu diesen Handbüchern

Keine Kommentare